4.4. Transfer the Drill to the Cover
Sterile Team Member
1. Present the Cover to the non-sterile team
member.
Nonsterile Team Member
1. Hold drill by battery and pay attention to
the orientation of the drill.
2. Drop the drill inside the Cover so that
it only touches the fabric on the inside.
The inside of the Cover is now
contaminated.
There is a risk of contaminating the sterile
team member's gloves during the transfer of
the drill. Ensure that the transfer only occurs
after the linen has been rolled back and the
sterile team member's gloves are protected
by the cuff.
Ensure that the sterile team member is
prepared to accept the drill without dropping
the sterile Cover.