5.4.2. Prevac (Dynamic Air Removal) Sterilization
1. Fold DrillCover linen once lengthwise and insert into a sterilization case (3/4 Aesculap
case or equivalent).
2. Separate the metal components from the cover by using a separate sterilization
container or partitions within the same container.
3. Sterilize per parameter set 1 or 2.
Parameter Set 1:
TEMPERATURE
EXPOSURE TIME
DRY TIME

6. DISPOSAL

DrillCover linen and hardware can be disposed with other biohazardous waste or it can be
sterilized and disposed as regular waste. Drill can be disposed of in common electronics
recycling.

7. STORAGE

Store device in a cool, dry place to maximize the lifetime of the product.
132ºC
4 minutes
30 minutes
Parameter Set 2:
TEMPERATURE
EXPOSURE TIME
DRY TIME
DrillCover Hex System: Instructions For Use
134ºC
3 minutes
30 minutes