Cres Cor CGMH73A 운영 및 유지보수 지침 매뉴얼
{카테고리_이름} Cres Cor CGMH73A에 대한 운영 및 유지보수 지침 매뉴얼을 온라인으로 검색하거나 PDF를 다운로드하세요. Cres Cor CGMH73A 6 페이지. Mobile decontamination cabinet
5925 Heisley Road • Mentor, OH 44060-1833
MOBILE DECONTAMINATION CABINET
OPERATING and MAINTENANCE INSTRUCTIONS
Model no.
CGMH73A
HOW INSTALL UNIT:
1. Place cabinet on level floor, in well ventilated area.
2. Unpack and install parts for hangers or racks.
3. Plug cord into wall outlet.
HOW TO RUN THE DECONTAMINATION CYCLE:
1. Read all instructions before operating the Decontamination Cabinet.
2. Turn on the device by pressing the ON-OFF Switch. The switch light will come on.
3. The unit starts heating to the previous cycle settings. This is standby mode. The cycle light will be off.
The Temp Control will display "LOTPr" until 140°F is reached. Then the setpoint will be displayed.
4. To begin the decontamination cycle, press the "CYCLE" button. The display will show cycle "rcP01"
and the "CYCLE LIGHT" will be flashing. Cycle "rcP01" setpoints are 185°F at 65% humidity for
15 minutes.
5. Press the DIAL to set the cycle.
6. The unit will start preheating and the Temp Control will display "PrHEt" until the cabinet temperature setpoint is
reached. Then "PUSH StArt" will be displayed. Do not press the start button at this time.
7. The cycle conditions are preset by the factory and are locked to prevent modification and ensure repeatability.
8. Do not run the unit if the low water indicator light is ON. The light will turn off once water is refilled.
9. Use the "DISPLAY" button to check for actual temperature and humidity as described under the
"Digital Control Notes" section on page 2.
Volts
Watts
120
2000
Phone: 440-350-1100 Ext: 268 • www.crescor-hs.com
Amps
Hertz
16.7
60
FIRST TIME OPERATION:
1. Wipe the inside clean with detergent and hot water.
2. Fill water pan with 4 gallons of hot water.
MOBILE
DECONTAMINATION
Rev. 0 (7/21)
Page 1 of 6
Model #CGMH73A
Phase
NEMA
1
5-20P
FL-2424