DIGIOXI PO-930 운영자 매뉴얼 - 페이지 11

{카테고리_이름} DIGIOXI PO-930에 대한 운영자 매뉴얼을 온라인으로 검색하거나 PDF를 다운로드하세요. DIGIOXI PO-930 19 페이지. Pulse oximeter monitor

"CHECK SITE", "WEAK PULSE", "LOW BAT"
04 - Indicates "ALARMS OFF" ( No alarm sound)
A

Initial Operation

.
CAUTION: Read this manual completely before initiate the monitor's installation and operation.
1 - Plug the AC power cord in AC MAINS receptacle ( 11 ) in rear panel and in the AC OUTLET. Verify if the
mains voltage corresponds to the voltage specified in the Rear Panel.
2 - The electrical installation of the room must have three pins outlets with earth connection for better protection
of patients, users and equipment's.
3 - Press the switch ( 02 ) in front panel to power-on the monitor.
4 - The monitor begins a self-test and displays the monitorization screen.
5 - At power-on the monitor assume standards values for the following items:
- Audible Alarm
- Pulse Rate High Limit
- Pulse Rate Low Limit
- SpO2 High Limit
- SpO2 Low Limit
B

Alarm Sound On-Off

.
Pressing one time the key (ALARM SOUND ON-OFF)
ALARM OFF message flashes in screen.
Pressing again the (ALARM SOUND ON-OFF)
SOUND ON-OFF) key for 4 seconds, disables permanently the ALARM SOUND and the message ALARM OFF
appear in screen permanently.
C

Alarm Limits Selection

.
At power-on the SpO2 and pulse alarm limits are pre-selected.
To modify the limits, press the alarm select (05) push-button until the desired limit ( SpO2 High, SpO2 Low,
Pulse High, Pulse Low ) flashes. Press the
select (05) push-button until no alarm limit indicator is flashing.
Operators Manual Pulse Oximeter Monitor DIGIOXI "PO-930"

SECTION 4 - Operation Instructions

: Two minutes OFF
: 140 BPM
: 60 BPM
: OFF ( -- -- )
: 87%
%
disables the ALARM SOUND for 2 minutes and
%
key enables the ALARM SOUND. Pressing (ALARM
é
ê
or
push-button to select the limit value. Press again the alarm
Digicare Biomedical Technology Inc.
9