Functional check

Move the lifting column out. Press the emergency stop switch approximately in the
middle of the vertical lift.
- It must be no longer possible to move the lifting column.
Unlock the emergency stop switch again.
- The lifting column can be moved again.
Move the lifting column into the top end position.
- The lifting column motor must switch off automatically at the top point of the lift.
- No squeaking noises should occur during the movement of the lifting column.
Move the lifting column in completely.
- The lifting column motor must switch off before the lowest end position is reached and
an acoustic warning signal must be audible.
- No squeaking noises should occur during the movement of the lifting column.
Briefly release the button for lowering the lifting column and then press it again.
- The lifting column can be moved into the lowest position. An acoustic warning signal
sounds during the movement into the lowest end position.

Protective conductor test

Perform the protective conductor test according to ARTD-002-731.17....
Observe the instructions for the protective conductor test in the first chapter.
SIREMOBIL Compact L
System Manual
The protective conductor resistance must not exceed 0.2 ohms.