Siemens SIREMOBIL Iso-C SP 설치 및 설정 지침 매뉴얼 - 페이지 15

{카테고리_이름} Siemens SIREMOBIL Iso-C SP에 대한 설치 및 설정 지침 매뉴얼을 온라인으로 검색하거나 PDF를 다운로드하세요. Siemens SIREMOBIL Iso-C SP 16 페이지.

Siemens SIREMOBIL Iso-C SP 설치 및 설정 지침 매뉴얼
Area dose product measurement device
Preparations
Attach and center the center cross on the I.I. (Fig. 4).
Select survey format.
Select KV mA curve plateau HC1.
Attach and center the small (1cm
measurement device (Dali, Nomex ...) or the semiconductor detector (Diados) onto the
image intensifier (Fig. 5).
Release fluoroscopy briefly, adjusting the iris on the radiation field to a format of
approximately 15 cm.
NOTE
Do not make any further adjustments to the iris.
Record the diameter (Fig. 6) of the octagonal area of the radiation field.
Remove the center cross.
Select kV STOP and enter 70 kV.
Dose measurement
Press the >0< key twice to reset the area dose product on the Siremobil monitor to 0 .
Reset the dose measurement device (Dali, Nomex ...) to 0 .
Release fluoroscopy several seconds. The measured dose K
that the uppermost measurement range of the dose measurement device is utilized fully;
if necessary, reset the area dose product display on the Siremobil and the dose
measurement device and repeat the measurement with a different fluoroscopic time.
Record the measured dose.
Record the area dose product displayed on the Siremobil monitor (FDPa).
Calculating the area dose product
To calculate the measured area dose product :
(measured area dose product FDPg) = (measured dose K
Evaluation
Calculate the deviation between the measured area dose product and the displayed area
dose product:
The deviation must be < 0.3 .
Deviation = [(displayed area dose product FDPa) - (measured area dose product FDPg) ]
/ (measured area dose product FDPg)
For deviations > 0.3 , the area dose product measurement device must be replaced.
Siemens AG
Medical Engineering
3
) dose measurement chamber of the dose
The radiation field must cover the entire area of the dose mea-
surement chamber respectively the semiconductor detector of the
dose measurement device.
SPR2-230.033.04
Rev. 03
) * (diameter)
E
Page 15 of 16
06.05
CS PS 24
should be high enough
E
2
* 0.829
SIREMOBIL Iso-C
15
1
1
1
1