Diagnostics and troubleshooting
7.1 On site troubleshooting
● In remote installation, coil cable screen must be connected at both sensor and transmitter
(PE). Electrode cable screen only connected at sensor
● If the coil cabling is correctly connected and the coil is not interrupted or shorted, then the
Verificator may be defective. It will advise you of this by failing the magnetism test and
writing "Self Check Error" on the certificate.
Table 7-1
Description
Adaptor for IP67 transmitters
Adaptor for 19" transmitters
Important
Please check that the earth wire has been mounted to the connection box.
34
Spare part ordering
WARNING
Order number
FDK:083F4165
FDK:083F4166
SITRANS F M Verificator
Operating Instructions, 07/2019, A5E45134218-AB