Technical Data

Pollution degree
Overvoltage category
Fuses information
Supply voltage
Power consumption
Output
Enclosure rating
Temperature
Verification accuracy
PC software
CE-confirmity
Traceability
Recalibration
Tests/approvals
Limitation
The Verificator automatically detects and sets itself to the correct supply voltage. A 115/230 V
1)
transmitter cannot be verified with 11 to 24 V AC/DC supply.
EN 50082-1 when power supply is established through power cables connected to front of Verificator.
2)
Connection
Mains Power cables connected to the Verificator may not exceed 2.5 meter length and shall
be properly grounded.
SITRANS F M Verificator
Operating Instructions, 07/2019, A5E45134218-AB
1)
WARNING
2
II
315 mAT (for mains)
2.0 AT (for others)
● 11 to 30 V DC
● 11 to 24 V AC, 50/60 Hz
● 115/230 V AC, 50/60 Hz, +10% −15%
< 30VA, incl. MAG 5000/6000 transmitter
RS 232 galvanically isolated
IP20 open briefcase / IP42 closed briefcase
● Ambient: −5 to 45 °C
● Media: 0 to 60 °C
● During storage: −40 to 70 °C
● Sensor: Magnetic field properties < ±2%
● Insulation; No degradation on system accuracy
● Transmitter: No degradation on system accuracy
PC with at least 486 processor and Windows 95, 98, NT or XP
● Emission: EN 50081-1
● Immunity: EN 50082-2
Calibration at Siemens Hagenau (COFRAC acredited laborato‐
ries) traceable to national and international standards
Mandatory recalibration with one year interval to maintain tracea‐
bility
Water Research Centre, England
Cannot be used on sensors in Ex areas, max. liquid temperature
60 °C
2)
8