DANGER
Load cells are machine elements not constructed with the usual safety factors.
Therefore appropriate crash protection or catastrophe protection must be provided
which suit the danger potential.
CAUTION
For protection against undesirable electrical currents such as they occur when
welding or from lightning, the load cells must be bridged with a highly flexible
grounding cable (e.g. SIWAREX R grounding cable 7MH3 701-1AA1).
The substructure must withstand the planned loads. The application surface must
have a roughness of a max. 1.6 µm.
The values in the technical data must be observed for the ambient conditions.
Note
The load cell must be protected from direct sunlight. The permissible operating
temperature could otherwise be exceeded and thus the accuracy impaired.
SIWAREX R Load Cells Operating Manual