Inspection and maintenance
2.5
Maintenance
2.5.1
Vertical drive
PMP
Vertical cylinder
•
Remove the two screws (3/Fig. 3) on the vertical cylinder and lubricate with LONGTIME
PD2.
Use the grease gun (Item no. 71 09 010) with the pointed nozzle.
•
Reinstall both screws (3/Fig. 3) and tighten them.
PMP
Gas springs
•
Lightly lubricate the polished metal parts (4/Fig. 3) of the gas springs with OPTIMOL
OPTIFLEX ELM.
PMP
Chain guide
•
Lightly lubricate the chain guide (5/Fig. 3) where accessible with LONGTIME PD2.
PMP
Toothed rack
•
Lightly grease the toothed rack (6/Fig. 3)of the potentiometer drive (r102) with
LONGTIME PD2.
Siemens AG
Medical Engineering
Fig. 3
RLL5-310.101.02
Page 9 of 16
Rev. 02
01.97
TDSP 1
2 - 9
UROSKOP D