System covers
Do not use abrasive cleaners, (due to possible material incompatibility) organic cleaners
or cleaning agents containing organic substances (e.g. benzine, alcohol, spot remover).
Do not spray wash the system. Cleaning fluids must not be allowed to seep into the system.
Clean and reinstall the covers. Reconnect all protective conductors.
2.11

Concluding steps

PMF
Protective conductor test
Perform the final protective conductor test according to TI 236, section VII.
Maximum total resistance: 0.2 ohm.
UROSKOP D
Inspection and maintenance
RLL5-310.101.02
Page 16 of 16
Rev. 02
01.97
TDSP 1
Siemens AG
Medical Engineering