Section 6. Servicing and Maintenance
Sensor and Plumbing
In many cases, a sensor may become fouled with oil or other foreign matter. It is often
possible to restore the functionality of contaminated probes with cleaning. If the contaminant
is known, the SNAP plumbing should be filled with a fluid that is known to be a solvent to that
contaminant. Ethanol/alcohol are very powerful
solvents and are highly effective against some
contaminants.
After cleaning with solvents, the plumbing should be
completely dried or flushed with a volatile solvent to
assure that it is dry prior to reinstalling it. If this is
not done, contamination of the system may result.
Maintenance
Your vacuum instrument should give you many years
of trouble-free service. There are no regularly
scheduled maintenance intervals. If consistent
accuracy is required, it is recommended that the
SNAP and power supply be returned for an annual
calibration check.
Notes on calibration
There is inherent drift in all sensors. Repeatability is specified in the datasheet, but the
specification is typically a worst case scenario as drift is not easily predicted and depends on
the operating environment. Depending on your accuracy requirements, it makes sense to set up
a calibration interval to obtain as found data, and get a fresh calibration. Having this information
will allow you to determine the optimal calibration interval. The accepted interval is 1 year, but
depends on the accuracy you require and what you have defined in your standard operating
procedures (SOPs).
Note the sensors have excellent accuracy by themselves. Additional accuracy is gained by
calibrating the sensor controller (sensor is integral on SNAP) to the sensor. The sensor itself
cannot be calibrated, but the sensor-controller pair is. See our blog on sensor interchangeability
effects on accuracy for the Bullseye Precision Gauge Piezo for more information.
©