Factor – The Factor (F) is a value from 0.1000000 to 1.999999. When the Factor is displayed, the first
digit is flashing. Press Yes to accept its value and activate the next digit, or No to edit. When editing,
press No until the desired value appears, then press Yes to accept. Repeat until all digits have been
accepted. When the Factor flashes on the display, press Yes to accept or No to re-edit. (Example: F =
0.257206)
Multiplier (E) – The settings are, [E 0] (Fx1), [E 1] (Fx10), [E 2] (Fx100), [E 3] (Fx1000), [E-3] (F/1000),
[E-2] (F/100), and [E-1] (F/10). Press No to display the next setting, Yes to accept. (Example: E = 0).
Note: The multiplier selections are limited when the capacity in grams is exceeded.
LSD – The Least Significant Digit (LSD) is the number of displayed divisions (d) by which the weight is
incremented. The values are Set 1, SEt 2, SEt 5, SEt 10, SEt 100 o SEt 0.5. Press No to go to the next
setting, press Yes to accept. (Example LSD = 1d) Note: LSD options are limited if the readability in grams
is exceeded.
The example custom unit will display [0.26 C] when 1g is placed on the pan.