15. Programming errors
MESSAGE
DESCRIPTION
preC.
Calibration error
Err.pn
Calibration error
Er 11
Calibration error
Calibration error
Er 12
Er 37
Calibration error
Er 39
Instrument not configured
C . er . 36
Calibration error
C . er . 37
Calibration error
Err.mo
Weight unstable
adC.err
A/D converter error
Cel.err
Global load cell error
Er.Cel.1
...
Load cell error
Er.Cel.4
SOLUTION
First calibrate the zero point (zero), then proceed with the
sample weight acquisition (span) (ch. 9).
Check the connection of the load cell. Check that the cell
signal is stable, valid and greater than that of the previously
acquired point.
Increase the calibration weight.
Check that the signal coming from the cell increases upon the
increasing of the weight loaded on the scale.
Repeat the calibration, checking that the capacity
and division have been correctly set.
Transmitter needs to be configurated.
Check that the signal coming from the load cell is not negative.
Check that the signal coming from the load cell is not negative.
Check in
parameter that the signal is stable.
adC.v
If the connection of the cells is with 4 wires, check that the sense
jumpers are inserted.
Converter failure. Reboot the instrument.
Signal anomaly: check the load cells connection.
Signal anomaly: check the indicated load cell connection.
10
QSG_ENG_DGT4X