Alcohol Countermeasure Systems Alcolab Testpoint 사용 설명서 - 페이지 9
{카테고리_이름} Alcohol Countermeasure Systems Alcolab Testpoint에 대한 사용 설명서을 온라인으로 검색하거나 PDF를 다운로드하세요. Alcohol Countermeasure Systems Alcolab Testpoint 12 페이지. Breath alcohol tester
TROUBLESHOOTING
PASSWORD
Each access level Owner, Administrator and Operator is password protected.
If the Administrator or Operator forgets their password, or their password needs to be
reset, the Owner will need to delete and recreate profile(s).
If the Owner's password is forgotten or needs to be reset:
1. Logout and Exit the application
2. Head to Settings.
3. Select Applications.
4. Select ALCOLAB testpoint.
5. Clear Data.
NOTE: All data and records will be lost on the ALCOLAB testpoint.
POWER
When operating on the built-in battery alone, the ALCOLAB testpoint may lose power
or enter a power saving mode.
To resolve this issue, use the AC/DC adaptor provided with the device. Plug the device
into an outlet and access the power button located on the back, left-hand side of the
ALCOLAB testpoint, under the rotating round cover. Press and hold the power button
for 5 seconds. Within a few minutes, the application will relaunch automatically and
resume regular operations.
During startup the device may require a software update. Device updates are
manufacturer recommended. Ensure the device is plugged into the AC/DC adaptor
provided to avoid power loss during the update process. If an interruption occurs during
a software update, the device may become unresponsive. Follow the Power steps
located above to resolve this issue, if the issue persists contact ACS.
CALIBRATION
When calibration is required, the device may prompt the user with an expiration
warning. Contact ACS for further instructions.
INSTRUCTION MANUAL
4