TROUBLESHOOTING
NOTE: Refer to the ALCOSIM simulator Instruction Manual for a list of error codes that
may appear on the simulator display.
ADMIN MODE PASSWORD INPUT
The ALCOSCAN tester returns to a ready state if the password is not entered within
15 seconds. The tester will also return to a ready state after the password has been
entered incorrectly with a "three beeps" indication.
The ALCOSCAN tester will automatically shut down if left inactive for over 3 minutes.
LED COLOURS
If the Status LED blinks red and the blue and red Result LEDs blink, then the alcohol
concentration detected is outside of the acceptable range. Use only 20 mg% alcohol
reference solution. Keep the correct distance of 10 cm from the diffuser (use only ACS
p/n 58-000707 diffuser). Ensure proper setup and follow the procedure step by step.
If the Status LED blinks red and a result LED blinks green after taking a test, then the
calibration has been attempted in an environment outside the temperature range of
20°C to 30°C.
SERVICE
Any product being returned for service must be properly packaged and shipped
(prepaid) to a local authorized service provider.
For more information, contact an authorized service provider or visit acs-corp.com.