YOU MUST READ THIS ENTIRE MANUAL BEFORE YOU LEAVE TODAY!!!
3.3. Warm Up
The Determinator must warm to operating temperature before a sample is requested. Bringing a unit indoors in
a cold climate will reduce the required warm up time. The message displayed while heating shows 0% to
100% completion.
3.4 Camera (Optional)
Some State Laws require the use of Cameras with some Breath Alcohol Ignition Interlock Devices. This camera is contained within the Hand Held
portion of your device. State Law states that any attempt to block or obscure the camera will be considered a violation, and can be considered
grounds for extension, and or cancelation of your Driving Privilege. Attempts to obscure the camera will result in a failed attempt to start, delaying
you from starting your vehicle. With that in mind be cautious of how you hold the device while testing, be sure to hold it level with the ground,
camera up (any other way will the device will not accept your sample) do not place your hands around the device, DO cup it or hold it on the sides,
so as not to interfere with the built-in camera.
How to properly hold the device during test!
3.5 Provide Sample: Blow/Suck/Blow (BSB)
The Determinator requests a breath sample at turn-on and at random intervals while you drive.
Do not provide a sample until requested
Two beeps are sounded for attention before "Please Provide Sample" is seen/heard.
Breathe in deeply to fill lungs.
Provide the sample in three steps:
o STEP #1: Blow: short, sharp, pulse of air until you hear a beep.
o STEP #2: Suck: short, sharp, pulse of air until you hear a beep.
o STEP #3: Blow: long, continuous stream of air to empty the lungs. Keep the pressure above/below
the min/max set points. Three tones may be heard:
Low tone : blow harder
Medium tone : try to maintain this
High tone : blow softer
When the tone stops, stop blowing, sample is complete. Wait for the voice prompt to start the vehicle. DO
NOT attempt to start the vehicle until instructed to do so, attempting a start before being prompted may result in a violation.
Breath Sample Procedure:
3.5.1Sample Errors
If an error is detected while taking a sample, the Determinator says "INVALID SAMPLE" at the same time the display will scroll one of
the following messages to report the actual error and requests the user retry the sample. Possible errors include:
"Err: Pressure Not Stable" => Prior to Step #1 - do not blow into the unit until requested
"Err: Suck Not In Time" => Step #2 (short suck) sample not provided in allotted time
"Err: Blow Not In Time" => Step #3 (long blow) sample not provided in allotted time, or you may be blowing too hard for it to register
"Err: Low Sample Pressure" => Step #3 - pressure was too low during final breath sample
"Err: High Sample Pressure" => Step #3 - pressure was too high during final breath sample
"Err: Sample Temp" => Step #3 –temperature out of range, Do not leave the device in direct sunlight as it may overheat the device
"Err: Fuel Cell Stability" => At the start of a sample, the alcohol measurement device was not ready