REF: Druglizer LE5 Drug Tester User Manual V7.docx
Uncontrolled when printed
Firstly, instruct the subject to insert the
collector pad into their mouth then:
Rub on inside of both cheeks, 4 times each.
Collect saliva around mouth and under the
tongue.
Ensure the sponge is not chewed.
Instruct the subject to insert the Collector
into their mouth and rub the pad onto the
inside of both cheeks as well as under the
tongue and around the mouth.
Place collector on top of tongue, close
mouth, and tilt head in a chin to chest
motion
IMPORTANT: Druglizer LE5 training MUST
be undertaken in order to ascertain correct
saliva collection. Repeat collection steps
until the Collector indicator changes to a
BLUE that is consistent from edge to edge
When enough saliva is taken, the Collector
Indicator will change to a BLUE that is
consistent from edge to edge
For optimal hygiene – both in a test and for
a Druglizer LE5 – ensure the sampled
Collector surface is not pooled with
excessive saliva.