Alcolock V3 캘리브레이션 매뉴얼 - 페이지 7

{카테고리_이름} Alcolock V3에 대한 캘리브레이션 매뉴얼을 온라인으로 검색하거나 PDF를 다운로드하세요. Alcolock V3 8 페이지. Alcohol gas standard

Alcolock V3 캘리브레이션 매뉴얼

TROUBLESHOOTING

In the event of a calibration failure, first check the common causes of errors:
• Plastic tubing has been overused or has condensation present
• The alcohol gas standard value is not 130 PPM, or the value has been incorrectly entered in
the Select Value menu
• The alcohol gas standard is expired
• All connections to and from the gas valve are not secured properly
• The value entered in Altitude menu is incorrect
ALTITUDE REFERENCE TABLE
The altitude adjustment is in increments of 200 meters. Round the altitude up or down to the
closest value as follows:
CALIBRATION MANUAL
IF ALTITUDE IS
0 - 100 m
101 - 300 m
301 - 500 m
501 - 700 m
701 - 900 m
901 - 1100 m
1101 - 1300 m
1301 - 1500 m
THEN SET TO
0
200
400
600
800
1000
1200
1400
5