8. Maintenance

Recommended at least once per year.
Check that
General
The walker feels firm/free from play.
The walker does not rattle during manoeuvres.
The walker is level and all of its castors are in contact with
the floor.
The walker is not dirty.
Armrests
Arm rests are intact and clean.
Arm restwidening is functional.
Handtag
Grips are not damaged/dirty.
Handle adjustment is functional.
Battery
Check that the charging cable is connected to the control
box.
Check that the remote control is connected to the control
box.
Check that connections are made according to the wiring
diagram.
Check that the battery, actuators and control box are not
loose.
Height adjustment
Raising and lowering are functional.
The walker feels stable at maximum settings.
There is no play between the vertical frame and bottom
frame.
Height adjustment clamps lock.
The electric actuator fitting must be free from play. Height
adjustment via remote control is functional. The electric
actuator must run smoothly at a constant speed.
Frame parts
There is no mechanical damage.
There are no scratches.
End plugs/lower frame fitted.
Castors/brakes
Castors roll easily/tread not damaged.
The castors are tightly fastened to the lower frame.
Castor brakes functional on all castors.
Handbrake functional.
6
BUR E O R THO /SPACE
Action
Clean with lukewarm soap solution or alcohol-based
cleaning agent (no petroleum products).
Clean; replace if damaged.
Fit lock knobs/plates as required
Clean with lukewarm soap solution (no petroleum
products).
Replace handle grips.
Fit lock knobs/plates as required.
Fit new cable; this must always be connected to the control
box (IP65 classification).
Connect or fit new remote control as required.
Refer to the wiring diagram in the Technical manual.
Tighten; replace with new fasteners as required.
Fit new guide sleeves (in frame) or end plugs (in chrome
tubes) as required.
Tighten all lock bolts.
Replace clamps.
Tighten the attachment fitting concerned; replace bolts/
lock nuts as necessary.
Check that the battery/remote control/ actuator are
connected in accordance with the wiring diagram. Charge
the battery. Contact customer services.
If damage is present, contact GATE's customer
services.
Touch up as necessary.
Fit new end plugs.
Clean or replace castors (NOTE: Castors are
always fixed to the lower frame with thread locker
or lock nuts. The castors are sealed and we do not
recommend their dismantling; instead, replace the
whole castor assembly.
Tighten the castor bolt and fit locking nut or use
thread locker (depends on model).
Replace castor assembly.
Adjust the brake or fit new castors.