18. Information concerning EMC
The Easy hygiene chair has been function tested in both hospital and home care environments. The hygiene
chair's principal functions are its ability to be operated up/down, back/forward together with battery charging.
Tests have shown that as long as the interference the chair is exposed to is within the limits of this standard,
no failure of the primary functions can be demonstrated. Interference that is more severe than demonstrated in
these tests, may affect the hygiene chair's functions such that manoeuvring or battery charging fails.
The hygiene stool includes the following wiring:
•
AC cable to the charger, 3-wire, 2.0 m
•
DC cable between the charger and the remote control, 2 parts, 1.0 m
•
Signal cable between the control box and remote control, 2.5 m
WARNING: Always replace the hygiene chair's electrical components or wiring with original parts from
the supplier. Failure to do so may affect the product's EMC approval.
Avoid placing/using the chair closer than 0.35 m from other medical equipment and/or mobile telephones (or
other transmitting equipment). This is because there is then a risk that the product's main functions may not
work as intended or that the product may interfere with other medical equipment.