8. Troubleshooting when chair cannot be operated
NOTE: Chair cannot be operated over seat height 65cm if chair is tilted forward. Seat can also not be tilted
forward if seat height is higher than 65cm.
Before starting troubleshooting - always restart the battery (see below - hold the battery OFF)
button for about 5 seconds. Wait a minute. Activate the battery through its ON button).
1.
Check the battery charge via the re-
mote control's battery indicator (acti-
vate the remote control if necessary by
pressing any button).
1a.
Indicator shows green ==> Calibrate
actuator (see page 7).
1b.
Indicator shows red ==> Charge the
battery.
1c.
Indicator is off ==> Connect the
charger to the remote control. If charg-
ing is indicated in the remote control,
charge the battery as instructed.
If charging is not indicated - continue
to 2.
If the above has been done and the chair cannot be operated - contact Customer Service / GATE.
9. Safety instructions
Shower chair should only be used:
•
As a shower / hygiene tool
•
Indoors, on flat and hard surfaces
10
EASY SO FTB ACK
2
Switch off the battery manually using
the Battery Off button (see below).
Wait a minute. To activate the battery,
press its On button (see below).
Make sure the battery indicator light is
on or flashing.
The indicator light is on / flashing ==>
Replace the remote control and charge
the battery as instructed.
The indicator light is not on / flashing
==> Change battery.