1. Introduction

The Mattress consists of a foam head cell and series of 14 transverse air cells, each containing a unique
foam profiled insert, which are in turn held within a foam U Core, all protected by a vapour permeable
waterproof cover. The single head end cell and the formers consist of foam only. The transverse cells are
arranged into alternate pairs of A and B cells which are filled and emptied in sequence.
In Static Mode, the mattress attains the pressure reducing properties of the Dyna-Form Mercury
static foam mattress (details available on request), whilst in Alternating Mode the mattress is able to
offer similar properties to a pressure relieving dynamic system.
The digitally controlled Power Unit controls a pump that allows air to flow into, or out of the air cells
as required according to the operating mode selected. It also maintains the air pressure within the
mattress at the required level and controls the action of the audible/visual Audible Warning System in
the event of mains supply failure or over or under inflation pressure. A CPR Valve located at the pump
end of the umbilical hose permits the rapid deflation of the Mattress in an emergency.
2. Quick Reference Guide
This is a quick reference guide for the Dyna-Form Mercury Advance Trolley System
Product Code MAT/ADV/TROL/SIZE
Power Switch Audible Warning Reset
The power switch simply switches the mains power to the pump on and off.
When the pump detects an Audible Warning condition, this can be silenced (see page 5)
and re-set by switching the pump off and then back on again.
CPR Valve
Please ensure that the CPR connector is always placed fully home, prior to inflating the mattress.
NB: The mattress will NOT inflate properly should this not be the case.
The CPR connector is only to be used in the event of a clinical emergency for priority use.
However, disconnecting this function will cleverly deflate air rapidly from the mattress in
readiness for transport / static mode.
4
D Y N A - F O R M  M E R C U R Y A D V A N C E T R O L L E Y
(Frequently used functions)
D I R E C T H E A L T H C A R E G R O U P . C O M