D Y N A - F O R M M E R C U R Y A D V A N C E
12. Technical Specification
Declaration – electromagnetic emissions - for all ME EQUIPMENT and ME SYSTEMS
Guidance and manufacturer's declaration – electromagnetic emission
The MAT1210001 is intended for use in the electromagnetic environment specified below.
The customer or the user of the system should ensure that it is used in such an environment.
Emission test
Compliance
RF emissions
Group 1
CISPR 11
RF emission
Class B
CISPR 11
Harmonic emissions
Class A
IEC 61000-3-2
Voltage fluctuations/flicker
Complies
emissions
IEC 61000-3-3
1 8
D I R E C T H E A L T H C A R E G R O U P . C O M
Electromagnetic environment – guidance
The system uses RF energy only for its internal function.
Therefore, its RF emissions are very low and are not likely
to cause any interference in nearby electronic equipment.
The system is suitable for use in all establishments, including
domestic establishments and those directly connected to
the public low-voltage power supply network that supplies
buildings used for domestic purposes.
S E R V I C E M A N U A L
12. Technical Specification
Declaration – electromagnetic immunity
Guidance and manufacturer's declaration – electromagnetic immunity
The MAT1210001 is intended for use in the electromagnetic environment specified below.
The customer or the user of the system should ensure that it is used in such an environment.
Immunity test
IEC 60601
test level
Electrostatic discharge (ESD)
±6 kV contact
IEC 61000-4-2
±8 kV air
Electrical fast transient/burst
±2 kV for power
IEC 61000-4-4
supply lines
±1 kV for input/output line(s)
Surge Immunity Test
± 1 kV line(s) to line(s)
IEC 61000-4-5
Voltage dips, short
Voltage Dips
Period
interruptions and voltage
%U
(Cycles)
T
variations on power supply
30
25
input lines
60
5
IEC 61000-4-11
>95
0.5
Voltage
Seconds
Interruption
% U
T
>95
5
Power frequency (50Hz)
3 A/m
3 A/m
magnetic field
IEC 61000-4-8
NOTE U
is the a.c. mains voltage prior to application of the test level.
T
D I R E C T H E A L T H C A R E G R O U P . C O M
Compliance level
Electromagnetic
environment – guidance
±6 kV contact
Floors should be wood,
concrete or ceramic tile.
±8 kV air
If floor are covered with
synthetic material, the relative
humidity should be at least
30%.
±2 kV for power
Mains power quality
should be that of a typical
supply lines
commercial or hospital
±1 kV for input/output line(s)
environment.
±1 kV differential mode
Mains power quality
should be that of a typical
commercial or hospital
environment.
Voltage Dips
Period
Mains power quality
%U
(Cycles)
should be that of a typical
T
commercial or hospital
30
25
environment. If the user of
60
5
the Span system requires
continued operation during
>95
0.5
power mains interruptions,
it is recommended that the
Voltage
Seconds
system be powered from an
uninterruptible power supply
Interruption
or a battery.
% U
T
>95
5
Power frequency magnetic
fields should be at levels
characteristic of a typical
location in a typical
commercial or hospital
environment.