the original calibration table into the FCORR.BAK file and lets you restore the factory calibration at a later
time.
Never change or delete the 'H*.TXT' / 'L*.TXT' calibration files on the power sensor. If these files are
damaged, the sensor must be returned to SatService GmbH and calibrated again. There is no way to
calibrate the power sensor on-site!