Nikon Eclipse Ti Series 사용 설명서 - 페이지 16

{카테고리_이름} Nikon Eclipse Ti Series에 대한 사용 설명서을 온라인으로 검색하거나 PDF를 다운로드하세요. Nikon Eclipse Ti Series 19 페이지. Differential i nterference contrast attachment

Chapter
4
Assembly
2.
3.
4.
5.
fl
Wnen using phase contrast microscopy
lf you also wish
to
perform phase contrast microscopy, install the phase
conlrast
obiectives and
the
condenser
cassettes
for
phase
contrasl
observation that correspond to the
condenser
lenses
to
be used, as
described
in
the
microscope instruction
manual.
Before beginning observation, be sure
to
center the
annular
diaphragm.
(Refer to the microscope instruction manual.)
The
assembly
procedure
is
now complete.
lf
you
require a very precise vibraîion direction adjustment, proceed with step
9.
lil
Optical System Direction Adiustment (Adiustment of vibration direction)
To assure
thal
DIC images can be observed at optimum contrast,
the
vibration direction of
lhe
polarizer and
analyzer must
be adjusted correctly.
1.
Focus and center
lhe
system condenser
as
describ€d
in
the microscope instruction
manual.
.
Bring
the A
(empty)
cassette
into
the optical
path.
When
using
the
condenser
slider,
remove the
condenser
cassette
that is
in
the
optical path so
the
slider
is empty.
.
Bemove the polarizer,
analyzer, and
objective
DIC prism
from the
optical path.
.
When
using
a
high
NA
condenser:
In
order
to
focus the
field
aperture
diaphragm
image on the
specimen surface,
and in order
to
avoid
striking lhe
spgcimen with the
condenser
lens, move
the condenser
lens
slightly
closer than
the
subiect distance
for each
condenser
lens, and
then
bring
the specimen
into focus
while
raising
the
condonser lens. When
using a
oil-immersion
type
condenser,
also
check
page
g.
Remove
the objective
DIC prism
that
ls located
directly
below
the
4Ox
obiective.
Swilch to
the
4Ox
objective and bring
the specimen
into focus.
Move
the stage
so that the
specimen
moves
oui
of the
viewfield.
In its
place,
bring an almost
dust-free
area
into
the viewfield.
(There musî
be
no dust in the whole
viewfield.)
Turn
the Bertrand
lens in/out lever
to the [Bl
position
to
bring
the Bertrand lens
into the optical
path
so
that the
obiective's
€xit pupil (the bright
circle)
can be
observed. (lf your
eyepiece
tube
has no Bertrand
lens in/out lever,
install
a
cenlering telescope
instead of an
eyepiece and rotate
its eyepiece part
to
focus
on
the obiective's
exit pupil.)
Vibration
dirgcton lgvgr
Vlbradon
dirgclion
lgver
damp scf€r',
(bottom)
Figurc
4-9
6.
When
using the
T-A2 analyzer, push
in
the
analyzer
to the
second clickstop so
that
it is in
the
optical path.
Loosen
the
clamp
screw
and align
the
index on the
vibration
direction
lever with
the
notch.
When using
the
epi-fl
filter turret
attached with the
analyzer
block, place
the analyzer block
into
the
optical path.
Teble
{-1
f
.gmm (including slide glass with
a
thid(rless of
f
.2mm)
15