II. ASSEMBLY
1/4
A
&
tint plate
Remove the screw by the side of
the 1/4
A
plate of the 1/4
A
& tint
plate and insert it into the com-
pensator slot of the intermediate
tube "P", facing the positioning
groove toward the operator side.
Reattach the removed screw.
®
®
CF Achromat
P objective
Mount the objectives on the nose-
piece in such positions that thfilir
magnifying power increases as the
nosepiece is revolved clockwise.
Input voltage change-over
switch
(For European districts only)
Make sure of the power source
voltage, 220V or 240V, by means
of the input voltage change-over
switch
on the bottom
of the
microscope base_
• To assemble the microscope, follow the procedure in the order given:
--'
CF eyepiece
_
®
Insert the eyepiece CFW 10XCM-
'"
into the right-hand sleeve, fitting
the pin of the eyepiece in the
right-hand
notch
of the sleeve.
Into the left-hand sleeve, insert
the CFW lOx.
Bottom of the Base
CD
@2
Achromat
strain-free condenser
Insert the condenser into
the con-
denser carrier,
fad ng the aperture
number
plate
toward
the operator.
Fasten the clamp screw on the left
side of the carrier.
Aperture
number plate
•. ----
Specimen clip
Place the clip on the stage using
holes on the stage surface.
Stage clamp screw
Circular graduated stage
Releasing the stage clamp screw
sufficiently,
mount the stage on
the circular dovetail of the sub-
stage. Clamp the screw.
Dia-polarizer
After centering the objec-
tives and condenser, insert
the dia-polarizer
into the
bottom of the condenser.
@
@)
@
Leveling foot screw
For
stable
installation
of
the
microscope, manipulate the adjust-
ing screw at one foot
on the
bottom of the microscope base.
6