BVI 9400 Calibration Quick Reference
BVI 9400 Calibration Quick Reference
The entire calibration process may take up to 15 minutes. You may calibrate your instrument as often as you like; you do not have
to wait for the next scheduled calibration date. However, at a minimum, you must calibrate the instrument every 12 months.
www.verathon.com https://my.scanpoint.com
For more information, call 800.331.2313 (in the U.S. and Canada) or contact your local Verathon Medical
Verathon Medical
Prepare for Calibration
Set up the calibration tank in a location that is within
10 feet of Charger and Wireless Hub.
Place the calibration tank on a flat, nonreflective surface and
remove the lid. Pour clean, room temperature water into the
container, filling to the indicator mark. Make sure the water
contains a minimal amount of bubbles. Place the spiral-
shaped target in the container, using the notches to position
target correctly. Replace the lid on the calibration container.
Place the probe
Place the Probe in the cutout in the top of the calibration
tank. Make sure the black tip of the Probe is submerged
in the water.
Open ScanPoint
QuickPrint
®
From your desktop, open QuickPrint by double-clicking
the QuickPrint icon.
Establish BVI 9400 - QuickPrint
connection
Press the ScanPoint
button on the BVI 9400.
®
On the QuickPrint screen, click the FIND NEW button.
When QuickPrint establishes a connection with the
BVI 9400, an icon will appear along the left menu.
®