3. Ensure that the power switch on the rear panel is in the OFF position ("O" side depressed). Plug the
C13 end of the included AC Power Cable into the C14 panel connector on the AC Power Adapter,
then plug the other end into a nearby AC power outlet. Press the "I" side of the power switch to
power the printer ON.
4. Printer is ready for use. Before using, calibrate the nozzle to the Z zero position.
Zeroing the nozzle
Before printing, it is recommended that you zero the nozzle. Zeroing the nozzle ensures that the tip is always
set to a reference zero position. Since most tips have a different length you would want to ensure when a tip
is changed the reference zero point is calibrated. To set the reference zero, do the following:
1. Do NOT install the syringe as yet.
2. Insert the microSD card and run the code "NOZZLESETUP.gcode". This code will move the
nozzle to the print platform.
3. If you're printing into a petri dish add it to the platform, directly under the nozzle.
4. Using the hex tool, loosen the two hex screws on the nozzle. DO NOT remove the screws. HOLD
THE NOZZLE and DO NOT REMOVE from the heatsink. Two personnel may be required.