Abbott Amplatzer TorqVue 사용 지침
{카테고리_이름} Abbott Amplatzer TorqVue에 대한 사용 지침을 온라인으로 검색하거나 PDF를 다운로드하세요. Abbott Amplatzer TorqVue 5 페이지. Exchange systems
TorqVue™ Exchange Systems
Description
The AMPLATZER™ TorqVue™ Exchange Systems consist of a delivery sheath, dilator, loader, hemostasis valve, delivery
cable, and plastic vise. The system components are identical to the AMPLATZER™ TorqVue™ Delivery System except that the
dilator has a larger inner lumen to allow passage over a delivery cable. The body of the delivery sheath is radiopaque for
visibility under fluoroscopy.
Figure 1. AMPLATZER™ TorqVue™ Exchange System Components (45° exchange system shown)
A. Delivery sheath – Provides a pathway through which a device is delivered.
B. Loader – Introduces a device into the delivery sheath.
C. Hemostasis valve with extension tube and stopcock – Allows flushing of the exchange system and controls back-bleeding.
D. Delivery cable – Attaches to an in vivo delivery cable to facilitate removal of a delivery sheath from a patient. Attaches to a device
for controlling its movement through the delivery sheath.
E. Plastic vise – Attaches to the delivery cable, serving as a "handle" for detaching (unscrewing) the delivery cable from a device.
F. Dilator (not pictured) – Eases penetration of tissue.
Indications for Use
The AMPLATZER™ TorqVue™ Exchange System is intended for removal of a delivery sheath and subsequent exchange with a
delivery sheath of equal or larger diameter.
Contraindications
None known.
™ Indicates a trademark of the Abbott group of companies.
‡ Indicates a third party trademark, which is property of its respective owner.
Pat. http://www.abbott.com/patents
© 2018 Abbott. All Rights Reserved.
Instructions for Use
A
C
B
1
D
E
ARTEN600038208A
2018-08
Manufacturing facility:
Abbott Medical
5050 Nathan Lane North
Plymouth, MN
55442 USA
+1 855 478 5833