Operation manual
Figure 4-2. Switching OFF/HI positions
Figure 4-3. Pressing the CAL button
Step 3
Press the test button (Figure 4-4). The CAL graphic will disappear and one
segment will remain displayed. The instrument has now calibrated itself
against the clean sample and is ready to test the oil in your system. The
calibration data will be stored in the memory until a new calibration procedure
is performed.
7
Operating instructions