Pressure cannot be maintained when calibrating another pressure gauge.
 Air is drawn into pump because of insufficient oil in center reservoir. Unscrew Cap and check oil level in
reservoir. Add oil if necessary.
 Damaged cylinder adapter O-Ring. Unscrew the piston cylinder unit from cylinder base, replace O-Ring.
 Leakage through valve.
Excessive oil leakage at piston.
 Damaged cylinder O-Ring. Replace O-Ring.
 Oil too light. Check and remove fluid in reservoir if it is shock absorber or brake fluid. Replace with mineral
oil or sebacate.
 Worn piston & cylinder assembly. Replace with new Piston & Cylinder Unit.
Not sensitive and/or inaccurate operation during hydrostatic testing.
 Water emulsifying with tester oil.
 Best - Use synthetic oil instead of standard mineral oil. Satisfactory - Use oil water separator.
Remark: The specification in this user manual would be changed without notice.