circular record.

15. Charging and Battery Maintenance

Note! Only for recorders with rechargeable battery (A and S-type models)
The battery starts charging automatically when you connect the recorder to the USB-
port of the computer. Full charging takes 3 hours (while charging LED is flashing
gradually with the interval of 5 seconds). With maximum charging LED will continuously
and smoothly change its brightness. The user can recharge the rechargeable battery via
power source with standard USB- output (5 V), during which the LED is permanently on
and one can assess the process according to the charging time only (3 hours).
Do not leave the rechargeable battery discharged. Recharge it at least once a month,
even if you don't use the recorder. Failure of the battery due to customer's fault is not
covered by the warranty.
Attention!
In firmware versions 1.14 for Tiny+ and 1.15 for Tiny16+ after memory filling, recorder
does not turn to Standby mode. That may lead to deep battery discharge and its
damage.
See your firmware version in EM PLUS program on the bottom line of the main menu
(for instance FW: 1.14).
In case of battery discharge, fulfill the following steps as soon as possible:
1.
Clear the memory
2.
Charge the battery
If you could not charge the battery, follow these steps:
1.
Connect USB adapter to PC
2.
Turn the switch to On position