Fitness Technology 400S+ 매뉴얼 - 페이지 4
{카테고리_이름} Fitness Technology 400S+에 대한 매뉴얼을 온라인으로 검색하거나 PDF를 다운로드하세요. Fitness Technology 400S+ 10 페이지. Performance force plate
PCB:
Self-contained in the 400S Platform Assembly.
FORCE PLATE 400 Kg Capacity
Load Cell Wiring
The four Load Cells connect into the XPV7 Electronics Module are colour coded on the
load cell cables at the terminal blocks for ease of wiring, they are as follows
Load Cell
1.
Load Cell
2.
Load Cell
3.
Load Cell
4.
The wiring connection for each individual load cell is as follows.
Sig + White
Sig - Green
Ex- Black
Shield (Heat shrink - wire colour may vary)
Fitness Technology, 21 Bishop Street, Skye SA 5072 Australia
Postal Address: PO Box 139 Fullarton 5063 Tel: +61 8 8331 9229
Copyright: Fitness Technology, Adelaide, Australia 3rd Feb 2020 Issue 1.00, 3rd Feb 2020
FORCE PLATE 1000 Kg
Capacity
Blue
Heat - Shrink shroud as per next page
Green
Heat - Shrink shroud as per next page
Red
Heat - Shrink shroud as per next page
Yellow
Heat - Shrink shroud as per next page
Ex+ RED
www.fittech.com.au E [email protected]
Terminal
Block is not
on 400 S Plate
Mob: 0418 815 400.
3