Artiss Easyspray 빠른 참조 매뉴얼
{카테고리_이름} Artiss Easyspray에 대한 빠른 참조 매뉴얼을 온라인으로 검색하거나 PDF를 다운로드하세요. Artiss Easyspray 2 페이지.
easysPRay Quick Reference Guide (for open surgery)
Instructions for Circulating Nurse | EasySpray Pressure Regulator
1
2
Insert 9V battery into the EASYSPRAY
Connect EASYSPRAY device to IV pole or
pressure regulator device.
trolley rail using the clamps on the back
of the device.
Instructions for Scrub Nurse | Spray Set
1
2
Prepare ARTISS Solution for Sealants
Firmly attach the spray head to the
according to the instructions in the
nozzle of the syringes.
package insert.
Instructions for Surgeon
1
2
Confirm (verbally) the actual pressure
Spray from a distance of 10 – 15 cm for
with theatre personnel.
optimum results.
3
4
Use suitable connection tube to connect
Connect Spray Set filters to EASYSPRAY
the EASYSPRAY device to medical air
device. Connect the blue filter to the blue
(ranging 3.5 – 7 bar / 50 – 100 psi).
female luer connector and the clear filter to
the male luer connector.
3
4
Fasten the pull strap to the double
Fit the connection tube of the spray set to
syringe system to assure the spray head
the luer-lock connector on the underside of
is tightly secured.
the spray head.
3
To activate the flow of gas occlude the
opening in the clip center with thumb.
To begin application, gently depress the
syringe plunger.
5
Turn the on/off switch on the front side
of the EASYSPRAY to the ON (I) position.
5
Attach the clip (on the end of the sensor
line) by sliding it into the grooves located
on the top of the syringe plunger.
The use of ARTISS Solutions for Sealant is restricted to experienced surgeons who have been trained in the use of ARTISS.
In order to ensure optimal safe use of ARTISS by spray application, apply a minimum spray distance of 10 cm and a maximum spray
pressure of 2 bar to minimise the potential risk of air or gas embolism, tissue rupture, or air or gas entrapment with compression.
The EASYSPRAY device will continue to emit gas for a brief period after the thumb is removed from the clip/plunger.
This delay helps to avoid clogging of the spray head.
Please see the ARTISS SmPC provided together with this material.
Caution: Any application of pressurised gas may be associated with a potential risk of air or gas embolism, tissue rupture or or air or gas
entrapment with compression, which may be life threatening. Be sure to take appropriate measures to address these risks by observing the
recommended minimum spraying distance and the maximum pressure provided in the appropriate spray set instructions for use.
6
Check the gauge on the EASYSPRAY
device for the appropriate pressure
range of 1.5-2.0 bars (21.5-28.5 psi).
Adjust pressure setting by turning the
black pressure control knob.
6
Pass the assembled applicator to the sur-
geon for spray application. Pass the end
of the connection tube with the sterile
filters to the circulating nurse.