DLab MS7-H550-S 사용자 설명서 - 페이지 13
{카테고리_이름} DLab MS7-H550-S에 대한 사용자 설명서을 온라인으로 검색하거나 PDF를 다운로드하세요. DLab MS7-H550-S 19 페이지. 7’square (hotplate) magnetic stirrer
7" Square Hotplate Magnetic Stirrer
transmission in start-stop-operation.
•
Mode of transmission: Fully Duplex. 1 start bit; 7
character bits; 1 parity bit [straight (even]); 1 stop bit.
•
Transmission speed: 9600 bit/s
•
Start remote control knob, LCD displays " Remote ".
Note:
Forbid to insert or remove the RS232C
communication line when switch ON!
9. Faults
•
Instruments can't be power ON
-
Check whether the power line is unplugged
-
Check whether the fuse is broken or loose
•
Fault in power ON self test
-
Switch OFF the unit, then switch ON and reset the
instruments to factory default setting.
•
Stir speed cannot reach set point
-
Excessive medium viscosity may cause abnormal
speed reduction of the motor
•
Unit cannot be powered OFF when switched OFF.
-
Check if the residual heat warning function is still ON
and hotplate temperature is above 50 ℃
LCD/LED screen still work and "Hot" flash).
If
these
faults
manufacturer/supplier.
10. Maintenance and Cleaning
•
Proper maintenance can keep instruments working
properly and lengthen its lifetime.
•
Do not spray cleanser into the instrument when
cleaning.
•
Unplug the power line when cleaning.
•
Only use recommended cleansers:
Dyes
Construction materials
Cosmetics
Foodstuffs
Fuels
•
Wear the proper protective gloves during cleaning of
the instrument.
•
Before
using
decontamination, the user must ascertain with the
(the
manufacturer that this method will not harm the
instrument.
11
are
not
resolved,
please
Isopropyl alcohol
Water containing tenside /
Isopropyl alcohol
Water containing tenside /
Isopropyl alcohol
Water containing tenside
Water containing tenside
other
method
for
cleaning
contact
or