Main interface is as figure 6-1. Left knob is setting speed, rotate the knob at the scale you want to set;
right knob is setting the time, rotate the time knob at the scale you want to set, then the centrifuge will run.
At the same time, the speed lamps will display when the speed arrives at the position.

6. Rotor Preparation

6.1 Prepare the samples

6.2 Inject the samples into tubes

CAUTION
Do not overload samples into the centrifuge which will cause leaking.
Do not exceed the actual capacity allowed in the user manual.

6.3 Keep the tubes balanced

Although the centrifuge can accept sample balancing by eye, we recommend that you keep this
centrifuge in a well-balanced condition to extend its life expectancy.
Never intentionally run the centrifuge under an unbalanced condition even though the allowable
imbalance is not exceeded.

6.4 Inspect the rotor

Check the rotor for corrosion or scratches before using.
CAUTION
If any abnormality such as scratches are found, stop using the rotor and contact our service center.
Only manufacturer's rotors must be used with the unit.

6.5 Symmetrically load centrifuge tubes into rotor

CAUTION
Make sure the rotor is securely fixed on the shaft. Otherwise, the rotor may be moved off while
rotating and cause damage to the centrifuge and rotor.
7