11.3.
Trigger Timing Chart
t1
Trigger input pulse width
From the start of the trigger
t2
Internal trigger signal rise
t3
Internal trigger signal pulse width
t4
Sensor pixel reset period
t5
Sensor exposure period
From exposure completion to data
t6
transfer
t7
Data transfer (sensor-FPGA)
t8
Data transfer (FPGA-USB → PC)
t9
Software processing time
t10
Valid trigger acceptance interval
10 μs or more
About 2 μs or less
About 10 μs (fixed value)
Shutter setting value * 1 line time
(1 line time = 1790 * 1/24 [μsec])
(1074 - Shutter setting value) * 1 line time
About 8 line hours
1024 line time ≒ 76 ms
It depends on the USB transfer speed.
(Approximately 40 msec at standard speed)
About 2 msec
(depending on CPU speed and processing content)
About 160 msec