ARTRAY ARTCAM-407UV-WOM-OP 지침 책자 - 페이지 16

{카테고리_이름} ARTRAY ARTCAM-407UV-WOM-OP에 대한 지침 책자을 온라인으로 검색하거나 PDF를 다운로드하세요. ARTRAY ARTCAM-407UV-WOM-OP 18 페이지. Usb2.0 ultraviolet ccd camera 1.5 m pixels
ARTRAY ARTCAM-407UV-WOM-OP에 대해서도 마찬가지입니다: 지침 책자 (19 페이지)

ARTRAY ARTCAM-407UV-WOM-OP 지침 책자
11.3.

Trigger Timing Chart

t1
Trigger input pulse width
From the start of the trigger
t2
Internal trigger signal rise
t3
Internal trigger signal pulse width
t4
Sensor pixel reset period
t5
Sensor exposure period
From exposure completion to data
t6
transfer
t7
Data transfer (sensor-FPGA)
t8
Data transfer (FPGA-USB → PC)
t9
Software processing time
t10
Valid trigger acceptance interval
10 μs or more
About 2 μs or less
About 10 μs (fixed value)
Shutter setting value * 1 line time
(1 line time = 1790 * 1/24 [μsec])
(1074 - Shutter setting value) * 1 line time
About 8 line hours
1024 line time ≒ 76 ms
It depends on the USB transfer speed.
(Approximately 40 msec at standard speed)
About 2 msec
(depending on CPU speed and processing content)
About 160 msec
- 16 -