2.3 Wiring
The pyrolyzer and fan
power connector
contains the yellow,
black, & white wires.
The pyrolyzer output
connector contains all
green wires.

Chapter 3 - Operation

Be sure both the pyrolyzer and the CL8 are powered on before operation. See the CL8 user's manual for
operation of the CL8 unit. No special configuration is necessary for the pyrolyzer option after the gas tables
are loaded at the factory.
3.1 Filters
End of line filters are recommended for all sample tubing on the CL8. See the CL8 user's manual for more
information.

3.2 Error Messages

The pyrolyzer option includes several event log messages in addition to the standard CL8 event messages –
see Figure 3 below.
Pyro Encl Fan Fault
Pyrolyzer fault Point X
***In addition to the enclosure exhaust fan, each pyrolyzer assembly has its own exhaust fan. When a
pyrolyzer fan fails all four points of the assembly will issue a fault message simultaneously. When the four
messages appear together check the pyrolyzer fan for proper operation.
DC-ITD-CL4-CL8MANPYRO_A
REV 8/25/2022
Connect the
pyrolyzer and fan
power connector
as shown.
Connect the
pyrolyzer
output/status
connector as
shown.
Message
Corrective Action
Repair/replace pyrolyzer enclosure exhaust fan.
Repair/replace the element for the specified point***
Figure 3
7