Etac immedia IM4110S 사용 설명서 지침 - 페이지 8
{카테고리_이름} Etac immedia IM4110S에 대한 사용 설명서 지침을 온라인으로 검색하거나 PDF를 다운로드하세요. Etac immedia IM4110S 9 페이지.
Etac immedia IM4110S에 대해서도 마찬가지입니다: 짧은 지침 (4 페이지), 짧은 지침 (4 페이지)
Thank you for choosing an Etac product. If the customer makes adjustments, repairs or combinations not predetermined by Etac, the
Etac CE certification and Etac warranty will be voided.
Warranty: Two-year warranty on material and manufacturing defects, provided that the product is used correctly. Before taking a
product in service always make a risk assesment according to national and/or local legislation.
First inspection:
•
Is the packaging intact?
•
Read the label on the packaging and check the article no. and product description
•
Check that Short Instruction is enclosed – updated User Manual/Instruction for use can be downloaded from www.etac.com or contact your
Etac Customer Service or the local distributor.
•
Check the label on the product - does it include article no., product description, lot/batch no., cleaning instructions and supplier name?
Periodic Inspection:
•
Make sure that your Manual transfer system always is in a perfect condition.
•
After wash – always check materials, stitching, handles, buckles.
•
If the product shows signs of wear and tear, it must be removed from service immediately.
•
Lot no/Batch no:
x
Production year
Visual inspection/Check the product:
Ensure that the materials, seams, stitching,
handles, buckles are intact/faultless.
Mechanical load or stress/Test handles:
Draw hard in the handles in opposite directions
and control the material, seams and stitching.
Test buckles:
Lock the buckle and draw in opposite directions,
control the buckle and the stitching.
Check always:
Manual transfer aids with handles, straps and buckles:
Manual transfer aids with high or low friction:
If in doubt, please contact Etac Customer Service or local distributor for more information and guidance.
For more information about training and the use of these products, contact the Etac transfer experts – www.etac.com.
Class 1 - European Council
Directive MDD 93/42/EEC
on medical devices
Service Information
x
x
x
x
x
Running number
x
x
x
x
Manual transfer aids for sitting and lying transfer:
The product can be
scrapped in accordance
with national regulations
8
Stability test:
Try to bend the product verify that it feels firm,
stable and solid.
Test low/high friction:
Place the product on a firm surface or a bed,
place your hands on the material and test the
friction by pushing your hands down into the
product.
Low friction – slides effortlessly
High friction – no sliding/moving
Read the
User Manual