Aseptico 120426 운영 및 유지 관리 매뉴얼
3
Loosely coil the motor cord when autoclaving. Avoid sharply
bending the cord.
4
Sterilize the motor and cord assembly using any one of the
STERILIZATION METHODS listed below ( page 15).
Syringe Tips
Syringe tips are autoclavable using any one of the STERILIZATION
METHODS listed below (page 15). Wipe off the syringe hand-
piece and hose using a soft cloth dampened with a disinfectant
solution. Do not immerse the syringe handpiece in any liquid nor
spray any liquid directly on the handpiece.
Scaler
(AEU-350S ONLY)
The scaler handpiece cover and scaler tips are autoclavable using
any one of the STERILIZATION METHODS listed below (page 15).
Wipe off the scaler handpiece and hose using a soft cloth damp-
ened with a disinfectant solution. Do not immerse the scaler hand-
piece in any liquid nor spray any liquid directly on the handpiece.
Unit and Footswitch
Clean the unit and footswitch by wiping with a soft cloth moist-
ened with a mild detergent or 1:10 bleach solution (1-part bleach
to 9-parts water). Use of solvents or other solutions may cause
damage and void the warranty.
AEU-350 / AEU-350S
Wrapped Sterilization — Place in an appropriately sized steriliza-
tion pouch and seal.
Flash Sterilization — For immediate use only.
O-Rings
Periodically apply PTFE lubricant to all o-rings to keep the system
operating at peak efficiency. Damaged or worn o-rings can leak
and degrade performance; replace as needed. Bulkhead o-rings
are located in the fittings for the saliva ejector line 
B
and the electric motor cord 
motor o-rings 
Replacement Parts table on page 18 for o-ring part numbers.
Electric Motor LED
The LED lens
soft and can easily be damaged
by sharp objects. Dust or debris
may cause a significant decrease in
optical output. Clean the LED using
a lint-free swab and isopropyl
alcohol. Do not use other solvents
as they may damage the LED lens.
STERILIZATION METHODS:
A
Standard autoclaving (Gravity displacement
method)
Time: 15 min
Temperature: 132° C (270° F)
Dry time: 30 minutes
B
Pre-vacuum (Dynamic air removal method)
Time: 4 minutes
Temperature: 132° C (270° F)
Dry time: 40 minutes
C
Unwrapped standard autoclaving (Gravity
displacement method)
Time: 10 minutes
Temperature: 132° C (270° F)
Dry time: No dry time is required for flash
sterilization.
C
D
(x3) are located on the motor shaft. Refer to the
B
C
A
A
on the motor is
A
the HVE line 
(x3) on the bulkhead. Electric
D
A