CAUTIONS (Continued):
including the connection to the unit. A poor connection can
prevent accurate measurements.
7. Always take an x-ray to confirm the results.
8. Do not clean the main unit with Gluteraldhyde, Ethanol, or
similar chemicals to avoid discoloration and surface corro-
sion of the unit. If the unit contacts any of these chemicals,
wipe off immediately with a damp cloth.
9. Indications vary slightly depending on individual tooth char-
acteristics and canal shape; the dentist should apply clinical
judgment on each case.
TROUBLESHOOTING:
1. Nothing happens on the display after the main switch is
pressed.
a. The button should be held down firmly for two seconds to
turn the unit on.
b. Verify the battery is properly installed.
c. Verify the battery is fresh.
2. The power indicator turns red and the unit turns off after
a couple of minutes.
a. Verify the battery is fresh.
b. The first battery of self-tests has failed, wait 5 seconds,
turn the unit back on. If this does not solve the problem,
contact your local dealer or our office listed on the back
page of this manual
3. The power indicator turns yellow and the unit turns off
after a couple of minutes.
a. The second battery of self-tests has failed, wait 5 sec-
onds, turn the unit back on. If this does not solve the
problem, contact your local dealer or our office listed on
the back page of this manual
4. The power indicator is yellow and blinking.
a. Replace the battery.
5. The display gives unstable indications.
a. The file may be following the canal at a sharp angle.
When this happens, the display sometimes "jumps" to a
perforation indication. If this seems the case, retract the
file just enough so the perforation indication stops. This is
the working length.
7.