Aseptico AHP-85MB-X 운영 및 유지보수 매뉴얼 - 페이지 12

{카테고리_이름} Aseptico AHP-85MB-X에 대한 운영 및 유지보수 매뉴얼을 온라인으로 검색하거나 PDF를 다운로드하세요. Aseptico AHP-85MB-X 16 페이지. Dental implant systems
Aseptico AHP-85MB-X에 대해서도 마찬가지입니다: 사용 지침 (2 페이지)

Aseptico AHP-85MB-X 운영 및 유지보수 매뉴얼

MESSAGES

System Messages
1 .
When the system is powered on, the software part number
and version will be displayed briefly on the console screen.
Error Messages
NONE
Fault Messages
NONE

WASTE DISPOSAL

Dispose of all medical waste products, residues, etc. in accor-
dance with your local, state and country laws. Dispose of the
AEU-1070-70V/AEU-1070 and accessories at the end-of-life in accor-
dance with your local, state and country laws.

CONTRAINDICATIONS

None known

ADVERSE REACTIONS

None known
12
AEU-1070-70V / AEU-1070
Aseptico Inc. warrants this product against defects in material or
workmanship for a period of two (2) years, from date of original
invoice. Some handpieces are warranted for one year under the
same conditions. Other handpieces and expendable components,
such as air turbines and light bulbs, are covered by shorter warran-
ty periods, or have no warranty. Aseptico's sole obligation under
product warranty is (at its sole option and discretion) to repair or
replace any defective component or product in part or whole.
Aseptico Inc. shall be the sole arbiter of such action.
In the event of alleged defect under warranty, the purchaser is
to notify Aseptico's Customer Service Department promptly.
Customer Service will provide instructions, usually directing that
the product be returned for service. Shipment to Aseptico and the
cost thereof is always the responsibility of the purchaser.
Accidental misuse, inappropriate installation, or failure to perform
directed maintenance voids the warranty. Deliberately defacing,
modifying, or removing the serial number voids the warranty.
Aseptico Inc. does not assume, under this warranty, any risks or
liabilities arising from the clinical use of its products, whether or
not such use involves coincidental utilization of products manufac-
tured by others.

TROUBLESHOOTING

Console does not light up when turned on:
• Check power cord connection.
• Check fuse. If blown, see Fuse Replacement (page 13)
Console lights up when turned on, but handpiece does not
operate:
• Check motor plug connection.
• Check foot controller connection.
• Depress the main foot controller switch.
• Increase the speed setting.
• Increase the torque setting.
• Check that the connector pins in the plug socket are not bent.
• Check that the rotary tool is properly seated in the handpiece
and that the latch/collet is closed.
No water flow from pump to handpiece:
• Check that the water setting is 10% or above.
• Check that water container seal is completely punctured.
• Make sure the irrigation tubing is properly installed in pump
door and flow is in the correct direction.
Motor slowing down or sluggish:
• Check for dirty, under-lubricated handpiece.
• Verify that the ratio setting matches the handpiece.
• Check if handpiece lubricant is draining into motor.
Display error:
• Switch off power, wait 5 seconds, then switch back on to reset.
Pump leaking:
• Replace the peristaltic pump segment of the irrigation line
with a new one from the extra tube set provided.

WARRANTY