The pulse output versions are recommended for best performance.
The analogue 4-20mA versions must be connected to a measurement or display
instrumentation with a total load impedance less than100 ohms, including the wiring
between the meter and the user equipment.
Applications information
Installation position
The unit should not be placed on the suction side of a pump,
always on the outlet side. The unit may be supported by
pipework of hoses due to its low weight, but should not be used
to support pipework or be subjected to vibration.
Liquids
The measuring tube is manufactured from a food grade plastic,
so is safe for use with drinking water and other beverages. Most
oils and other non-corrosive liquids can also be used. Liquid
and ambient temperatures must not be exceeded.
Mechanical Installation
The
unit
may be installed at any angle, although it is recommended that it be
installed with the pipe entry and exit horizontal and the cable entry at the top, so
that the U-shaped measurement section remains full of liquid, should the process
empty of liquid.
3/8BSP plastic fitting must be used to connect to the UF25
damaging the pipe connection threads. DO NOT OVERTIGHTEN. Max tightening
torque 3Nm.
Suitable John Guest pushfit connectors are PI451613S for the UF25 and PI0412S
for the UF08B100
The outer case is sealed to IP66
Electrical Installation
An eight core cable is used to connect to this unit. This should be connected as follows:
Supply: - Red: + (8 - 2 Vdc only), Black: common ground, Green: ground (internally connected to black).
4
Selectable outputs:- Blue: NPN Pulse, White: PNP Pulse, Orange: Voltage 0-5Vdc, Brown: Current 4-20mA
Yellow wire should not be connected at all.
Cynergy3 Components Ltd.
7 Cobham Road
Ferndown Industrial Estate
Wimborne, Dorset BH21 7PE
Telephone +44 (0) 1202 897969
Email:[email protected]
9001
ISO
C E R T I F I E D
© 2014 Cynergy3 Components, All Rights Reserved. Specifications are subject to change without prior notice. Cynergy3 Components and the Cynergy3 Components logo are trademarks of Cynergy3 Components Limited.
unit
Please see connection circuits overleaf
Pipe inlet diameter
A pipe diameter that matches the inlet on the unit is
recommended. Inlet bore diameters are 10mm on the UF25 and
7mm on the UF08.
Entrained gas/solids
This unit is designed to operate with clean liquids only. If
amounts of gas or solids increase beyond 1% by volume,
accuracy can be affected. Higher levels of gas or solids can
prevent transmission of the ultrasonic signal and the output will
fall to zero.