Ashcroft 2036 사용 설명서 - 페이지 13
{카테고리_이름} Ashcroft 2036에 대한 사용 설명서을 온라인으로 검색하거나 PDF를 다운로드하세요. Ashcroft 2036 20 페이지. Digital sanitary gauge
Step 5: Press the ENTER key to finalize your choice. The
gauge display will now display SET. After two seconds the
screen will display the pressure value for 0% on the top
line. The middle line indicates the bar graph at 100% of
fullscale. The bottom line will display SET 0.
Step 6: Press the
key) on the keypad to increase or decrease gauge value at
0% of range.
Step 7: Press the ENTER key to finalize your choice. The
new values for the bar graph and 4/20mA settings have
now been saved.
OFF: This option sets the amount of time before the gauge will
turn itself off. Offerings are Never*, 30MIN,10MIN, 5MIN, 2
MIN.
To use the OFF option:
Step 1: Press the MENU key.
Step 2: Press the
key) until the word OFF appears.
Step 3: Press ENTER.
Step 4: Press the
key) to select the desired OFF time.
Step 5: Press the Enter key to finalize the OFF setting.
UPDATE: This option allows for changing the rate at which
pressure is updated on the display screen. This feature is
useful with rapid changes in process pressure that may
cause flutter of the display. Options are 100ms*, 1 sec,
500ms and 200ms, updates per second or 100ms*.
Since customer processes vary, update rates should be
selected based on the application.
To use the UPDATE option:
Step 1: Press the MENU key.
*Indicates default.
MENU FUNCTIONS
(up arrow key) or
(up arrow key) or
(up arrow key) or
– 13 –
(down arrow
(down arrow
(down arrow