LED/LIGHT PROTOCOL
LED
Green
Red
ON
Flash
ON
Flash
4 Maintenance
Clean the sensor with water and a brush without using:
• High-pressure cleaner
• Highly compressed air
• Solvents
• Corrosive/caustic agents
• Alcohol-based disinfectants
During cleaning and disinfection, the sensor should be removed and the sealing plug for the ca-
ble should be mounted. Alternatively, the sensor protective cap can be mounted. The sensor
must be placed with the tip pointing upwards during disinfection.
After the sensor has been exposed to water and condensation, it requires a period where the relative humidity
is lower than 80 %RH in order for it to measure correctly.
Do not bend the sensor as this would inflict permanent damage on the electronics of the sensor.
5 Technical data
Measurement range
Accuracy
Output signal
Time constant
Supply voltage
Current
Load
Recommended load
Max. output current
Output impedance
Cable
DOL 104 0-5 V
Status
Operation OK
Outside normal range
(below 10 % RH or exceeding 95 %
RH)
Connection error
Load < 500 Ω
Sensor defect
Over or under voltage alarm
Overload
% RH
% RH
V
T
63
V DC
mA
Ω
kΩ
mA
Ω
± 3% RH (40–85 %) ± 4% RH (10-95 %) at 0-40 °C
0.05 V / % RH
20 s at 0.5 m/s air speed
12 mA no load
35 mA max. load
> 500 Ω - < 10 MΩ
2 m 3 x 22 AWG / 0.34 mm
www.dol-sensors.com
Functional graph
0 - 100
11 – 30
≥ 100
20
< 1
2
Technical User Guide