ASHE STL-96 운영 매뉴얼 - 페이지 6
{카테고리_이름} ASHE STL-96에 대한 운영 매뉴얼을 온라인으로 검색하거나 PDF를 다운로드하세요. ASHE STL-96 17 페이지. Stream selector six channel
The front panel of the Stream Selector includes the following features:-
1) A 20 character x 4-linealphanumeric OLED Display (with backlighting)
2) Six nos. of touch-sensitive "Capsense" keys.
3) Six nos.
Orange
LEDS to indicate status of the Stream Outputs.
4) Six nos.
Red
LEDS to indicate status of the Low Alarms.
5) Six
nos. Red
LEDS to indicate status of the High Alarms.
6) Three nos.
Red
LEDS to show the status of the Auto, Manual & Pause modes
7) Two LEDs of Red & Green to show status of RUN (Green) and STOP (Red).
CONTROL KEYS
The instrument has sixtouch-sensitive "Capsense" keys on the front panel, the functions of which are
described below:
This key is used for all instrument settings and calibration, as follows:-
Input Calibration, Input Range, Skipping of Streams, Alarms Settings (Low Alarm & High
MODE
Alarm), Retransmission outputcalibration (4-20mA DC) and Operation Mode Setting.
In the "running" mode, this key is used to switch over to the next stream,
SHIFT
In the "setting" mode, this key is used to shift the cursor.
This key is an alternating key between PAGE/INCREMENT options.
PAGE
In the "running" mode, this key is used to show the input status of all Streams together.
INC
In the "setting" mode, this key is used to increment the value of the selected digit.
AUTO
This key is used to select alternatively between AUTO/MANUAL modes of operation.
MANUAL
This key is used to PAUSE the process. The microcontroller will stop the process,while
PAUSE
keeping the relay status unchanged. Also the same key is used to escape from
ESC
parameter setting mode.
START
This key is used to toggle the START and STOP process.
STOP
POWER ON SCREEN
The ASHE STL-96operates on universal AC power supply of 90 to 270VAC, 50Hz.
When the instrument is powered ON, the display will go blank with only backlight for a period of 5
seconds.
This time has been provided for power supply stabilizationwithin the instrument.
The Power ON Screen on the OLED Displayas indicated for five secondsshall be as follows –
and
SETTINGS OF PARAMETERS
ASHE
STREAM SELECTOR
MODEL NO. STL-96
ASHE CONTROLS PRIVATE LIMITED.
S
S
STL-96
TREAM
ELECTOR
Operation Manual