LEAKY PLASMA BAG

If a Plasma bag leaks, turn the unit off and discard the leaking bag. The remaining bags are not contaminated, because the
ports are always kept out of the water. The bath water is contaminated and must be drained and the bath cleaned. See

CLEANING the BATH.

CLEANING the BATH
For your protection use gloves when working with any blood products. Turn the unit off.
1. Remove the RACK and LINK and take it to the sink, clean under running water and wipe with bleach
towelettes or a 10% bleach solution. Rinse off in running water.
2. Drain the bath by attaching the provided tubing to the DRAIN valve. [ See page 9 ] Tilt the bath to drain
as much water as possible through the drain valve.
3. Wipe the BATH with bleach towelettes or a 10% bleach solution. Use only water and bleach towelettes or
a 10% bleach solution for cleaning. Use alcohol for removing oil or grease. DO NOT IMMERSE THE
WHOLE BATH. WIPE THE PUMP SCREEN FROM THE OUTSIDE ONLY. DO NOT REMOVE
THE SCREWS. Rinse out the bleach with water.
4. Remove the drain tubing and fitting to close the DRAIN valve. Fill the bath with water. If you use
deionized or distilled water add a little pinch (about 0.01 gr.) of table salt. .
#ALG240 (1.0 ml. per 4 liter of water. Available USA only) to the water. Any other bath clearing agent
can also be used.
5. Mount the RACK
Reassemble the unit. See SETTING UP.
4
Add Algaecide Part