Ambient Weather WS-DP02 사용자 설명서 - 페이지 5
{카테고리_이름} Ambient Weather WS-DP02에 대한 사용자 설명서을 온라인으로 검색하거나 PDF를 다운로드하세요. Ambient Weather WS-DP02 10 페이지. Digital psychrometer with temperature, humidity, heat index, dew point, wet bulb temperature and clock
2 Display Settings
Note: The console has three buttons for easy operation: SET button, + button, and ALARM
button (which also serves as the – button).
2.1 Set Mode
To set the time and units of measure, press and hold the SET button 3 seconds or longer, and let go of
the SET button.
1. 12/24 Hour. 12 Hr or 24 hour will begin flashing. Press the + button to switch between 12
hour and 24 hour display mode. Press the SET button (do not hold) to advance to the next
setting.
2. Hour of Day. The hour of day will begin flashing. Press the + button to increase the hour, and
the ALARM button to decrease the hour. Press the SET button (do not hold) to advance to
the next setting.
3. Minute of Day. The minute of day will begin flashing. Press the + button to increase the
minute, and the ALARM button to decrease the minute. Press the SET button (do not hold)
to advance to the next setting.
4. Temperature Units of Measure. The °F or °C icon will begin flashing. Press the + button to
switch between °F and °C. Press the SET button (do not hold) to exist the Set Mode.
2.2 Dew Point Alarm Set Mode
To set the Dew Point Alarm, press and hold the ALARM button for 3 seconds or greater, and the dew
point alarm set point will begin flashing. To increase the dew point alarm set point, press the +
button. To increase rapidly, press and hold the + button. To decrease the dew point alarm set point,
press the ALARM button. To decrease rapidly, press and hold the ALARM button.
To confirm the Alarm setting and return to normal mode, press the SET button (do not hold).
2.3 Dew Point Alarm
To turn on the Dew Point Alarm, press (do not hold) the ALARM button. The alarm icon
appear. To turn off the alarm, press the ALARM button again, and the alarm icon will disappear.
2.4 Temperature, Wet Bulb, Heat Index and Dew Point
Display
Press the SET button (do not hold) to switch the display between temperature, heat index and dew
point, or:
temperaturewet bulbheat indexdew point temperature
2.5 Calibration Adjustment Mode
The purpose of calibration is to fine tune or correct for any sensor error associated with the devices
margin of error. The measurement can be adjusted from the console to calibrate to a known source.
Calibration is only useful if you have a known calibrated source you can compare it against, and is
optional. This section discusses practices, procedures and sources for sensor calibration to reduce
manufacturing and degradation errors. Do not compare your readings obtained from sources such as
the internet, radio, television or newspapers. The purpose of the heat index monitor is to measure
conditions of your surroundings, which vary significantly from location to location.
Version 1.0
©Copyright 2012, Ambient LLC. All Rights Reserved.
will
Page 5