CO2 CALIBRATION 
The meter is calibrated at 400 ppm (parts‐ 
per‐million) CO2 concentration at the factory. 400 
ppm is comparable to fresh outdoor air.   
 
There are 2 types of calibration available: ABC 
(Automatic Baseline Calibration) or manual 
calibration. One or the other should be used 
regularly to maintain good accuracy. 
Note: 
If TIM10 accuracy becomes a concern after long 
use or other special conditions, return the meter 
to the dealer for calibration. 
CAUTION: 
Do not calibrate the meter with air or bottled gas 
that contains an unknown CO2 level. The meter 
will assume it is 400 ppm CO2 and will give 
inaccurate measurements. 
 
ABC Calibration 
ABC (Automatic Baseline Calibration) is designed 
to eliminate the zero drift over time of the infrared 
CO2 sensor. The ABC function is enabled by default 
from the factory. 
For ABC to work properly, the meter must 
continuously monitor CO2 levels for at least 7.5 
days (power on). During that time, it will record 
the lowest CO2 level and set that as the 400 ppm 
baseline for future readings. Therefore, it is critical 
when using ABC calibration that the meter be 
exposed to fresh air at least once every 7.5 days. If 
this is not possible, then manual calibration is 
recommended. 
 
Manual Calibration 
Manual calibration requires fresh, dry outdoor air 
when CO2 levels are approximately 400 ppm. Do 
not calibrate during rainy days. High humidity will 
affect the CO2 levels in air.   
 
During manual calibration, avoid placing the meter 
in areas crowded with people or close to 
structures that might give off CO2 like air 
ventilation pipes, chimneys, or automobiles. 
 
To manually calibrate the meter, place it outdoors, 
and turn it on for at least 2 minutes. Then hold 
down "SET" ,"▲" and "MIN/MAX▼" 
simultaneously more than 1 second to enter CO2 
calibration mode (Fig.19). 400 ppm and   
 
 
 
 
 
"CO2" will blink during calibration. 
Wait about 30 minutes until the blinking stops. 
Calibration is completed automatically. To stop 
manual calibration, press "RESET" for at least 1 
second. 
 
Note: 
Keep the meter away from any animals, humans or 
plants which might affect the CO2 concentration 
nearby during the calibration. 
 
RH CALIBRATION 
This humidity accuracy of this desktop CO2 meter 
can be re‐calibrated. 
 
Contact CO2Meter.com to purchase a calibration 
kit with instructions. 
 
SPECIFICATIONS 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
TROUBLESHOOTING 
? Cannot power on 
      Check whether the adaptor has power and 
      that the meter is firmly plugged in. 
? Slow response 
      Check that the air flow channels on 
      the rear of the meter are not blocked. 
? No change in readings 
      Check whether the meter is in maximum 
      or minimum mode. Reset factory defaults. 
? "BAt" and green LED keep flashing 
      The adaptor output voltage is too high or 
        too low. Use the supplied adaptor with   
        correct 5V (+10%) >=0.5A ratings. 
 
 
? Error Codes